ELBA - BACTERICIDAL lighting fixture type BSR 02 1X15W TUV 15 equipped (UV-C, 254 nm), without lamp(s)- for medical use
Technical characteristics
- Supply voltage: 230V/50Hz
- Ambient temperature: -5 ̊C – +35 ̊C
- Relative humidity: 80% at a temperature of +20 ̊C
- Saline fog: 48 hours according to SR EN 60068-2-11
- Shock resistance: 1000 ±10 shocks, 10g acceleration, pulse duration: 16ms, according to SR EN 60068-2-29
- The lighting fixture is made and validated in accordance with the requirements of SR EN 60598-1, SR EN 60598-2-1 standards.
Mounting
- The BSR-02 type lighting fixture is intended for disinfecting rooms mainly in the absence of people.
- Apparent installation on the wall above the door or mobile on a tripod directed towards the surface to be disinfected.
BSR-02-1xT8 2xT8 ORIENTABLE BACTERICIDAL TUV
- The BSR-02 type lighting fixture is intended for disinfecting rooms in hospitals, medicine factories, food factories, food warehouses, etc. mainly in the absence of people.
- The light fixture consists of a welded and painted steel sheet casing.
- Mirrored aluminum sheet reflector with reflection coefficient >0.85.
- The product is equipped with a joint that allows it to rotate around an axis parallel to the tube. This fact allows the beneficiary the possibility of directing the flow of UV radiation to certain areas in the room where the product is installed.
- The body support allows rotation in the vertical plane by ~140º. To allow horizontal orientation, the body is mounted on a tripod, which also ensures the possibility of approaching the surfaces that need to be disinfected.
- Apparatus (ballast) included in the product and executed in accordance with specific regulations.
Field of use
The BSR-02 type lighting fixture is intended for disinfecting rooms in hospitals, medicine factories, food factories, food warehouses, etc. mainly in the absence of people.
Bactericidal lighting fixtures disinfect the air and surfaces by irradiation with type C ultraviolet rays, destroying, neutralizing microbes and their multiplication.
UV radiation destroys: bacteria, viruses, fungi, molds, algae.
Disinfection of air, water and surfaces in hospitals, bacteriological research laboratories and pharmaceutical environments, food processing industry, food storage areas, food packaging, etc.
For disinfection of the air in the upper part of the room in hospitals and intensive care units and in the presence of people, products using UV-C radiation lamps must be oriented to radiate upwards and must be located so that the UV lamps do not reach the field at all visual. UV-C radiation from the lamps affects the eyes and skin.
Recommended light source
The product is delivered without lamp(s).
Short-length UV radiation lamps with a maximum at the UV-C 253.7nm wavelength for disinfection purposes: TUV T8 15W; TUV T8 30W, Dulie G13
T8 tube UV-C radiation power: TUV 15W 4.9W, TUV 30W 12W
The TUV T8 lamp is equipped with a warning sign indicating that the lamp emits UVC.
Coverage area and features may vary depending on the type and power of the lamp with which the product is equipped. Contact the lamp manufacturer for customized solutions. Indicative data:
1 x TUV 15W lamp: Intensity from a distance of 1m: 0.9 W/m2, effective surface 6-8 m2, effective sterilized volume 18m³. The sterilization capacity depends on the air circulation efficiency.
x TUV 30W lamp: Intensity from a distance of 1m: 2.3 W/m2, effective surface 12-15 m2, effective sterilized volume 36m³. The sterilization capacity depends on the air circulation efficiency.
2 x TUV 30W lamp: Intensity from a distance of 1m: 3.6 W/m2, effective surface 10-18 m2, effective sterilized volume 25-60m³. The sterilization capacity depends on the air circulation efficiency.
Documents & Downloads:
Technical sheet https://www.elba.ro/wp-content/uploads/BSR-02-115_130_230_Bactericid_Ed8sn.pdf
Instructions for mounting https://www.elba.ro/wp-content/uploads/BSR-02-115-O-BSR-02-130-O-IME.pdf
Response ANIMDMR https://www.elba.ro/wp-content/uploads/Raspuns-ANMDMR.png
Manufacturer | ELBA |
---|---|
Manufacturer code | 26372014 |
Order Unit | Pce |
Colour | White |
Weight Kg | 1,25 |
Mounting and installation | wall-mounted |
IP degree | IP20 |
Power | from 20 to 100 |